Selenium testing for total recoverable concentrations by conventional ICP-MS techniques is extremely prone to mass spectral interferences. The plasma gas (argon) and constituents of the sample matrix (calcium, carbon, chloride, sulfur, etc.) can easily combine to form polyatomic ions with the same mass-to-charge ratios as the various isotopes of selenium, resulting in false-positives and elevated detection limits.
The table below lists the different isotopes of selenium and the common interferences that can affect measurements at those isotopes.